[ Identification | Description | Input parameters | Links ]
ILL_H53_IN14
InstrumentIN14 is a high flux cold neutron three-axis spectrometer situated on a straight 58Ni coated guide looking at the horizontal cold source. IN14 is equipped with a vertically curved (002) pyrolytic graphite monochromator. A polarising bender can be inserted after the monochromator, for polarised neutron work. The sample, analyser and detector units are mounted on standard Tanzboden modules and the distance between these units may be changed if necessary. For sample positioning, two motorized arcs (+/- 20deg), and two horizontal translation tables (+/- 10 mm) are available. Soller collimators can be easily inserted before and after the sample, and before the detector. Three horizontally focusing analysers are available: (002) pyrolytic graphite, bent Si(111), and Heusler (111) for polarised neutrons. In this TAS configuration, PG002 is used as monochromator analyser, with a single type detector.
Name | Unit | Description | Default |
KI | Angs-1 | central wavevector for incoming neutrons | 1.55 |
QM | Angs-1 | wavevector transfer modulus at the sample | 1.0 |
EN | meV | energy transfer at the sample | 0.0 |
verbose | toggle verbose mode | 1 | |
WM | m | Width of monochromator | 0.15 |
HM | m | Height of monochromator | 0.12 |
NHM | 1 | Number of vertical slabs composing the monochromator | 1 |
NVM | 1 | Number of horizontal slabs composing the monochromator | 9 |
RMV | m | Monochromator vertical curvature, 0 for flat, -1 for automatic setting | -1 |
DM | 3.355 | ||
WA | m | Width of analyzer | 0.20 |
HA | m | Height of analyzer | 0.10 |
NHA | 1 | Number of horizontal slabs composing the analyzer | 11 |
NVA | 1 | Number of vertical slabs composing the analyzer | 1 |
RAV | m | Analyzer vertical curvature, 0 for flat, -1 for automatic setting | 0 |
RAH | m | Analyzer horizontal curvature, 0 for flat, -1 for automatic setting | -1 |
DA | AA | Analyzer lattice spacing | 3.355 |
SM | 1:left, -1:right | Scattering sense of beam from Monochromator | 1 |
SS | 1:left, -1:right | Scattering sense of beam from Sample | -1 |
SA | 1:left, -1:right | Scattering sense of beam from Analyzer | 1 |
L1 | m | Source-Monochromator distance. Contains 1st Collimator | 7.0 |
L2 | m | Monochromator-Sample distance. Contains 2nd Collimator | 2.12 |
L3 | m | Sample-Analyzer distance. Contains 3rd Collimator | 1.37 |
L4 | m | Analyzer-detector distance. Contains 4th Collimator | 0.7 |
ILL_H53_IN14.instr
.